22 / CAP TODAY February 2010
Bedside glucose testing systems
Nova Biomedical Roche Diagnostics
Sales Department [email protected]
Accu-Chek Customer Care Service Center
200 Prospect St. 9115 Hague Rd., Indianapolis, IN 46256
Waltham, MA 02454 800-440-3638
Part 6 of 7
781-894-0800 or 800-458-5813 www.novabiomedical.com www.roche-diagnostics.us
Name of instrument/First year sold
StatStrip Glucose Monitoring System/2006 AccuData GTS, 1994; AccuData GTS Plus, 2000
Professional or home use
professional use
professional use
List price
call for pricing, includes bar-code reader, spare battery, quick reference
guide
$550
Units sold in U.S./Outside U.S./In 2009
40,000*/5,000/—
Part of series of similar or related models
yes yes
Dimensions (H × W × D)/Weight
6.0 × 3.25 × 1.8 in/0.8 lb 11 × 8.75 × 4 in/5 lb
Analytical method/Technology/Enzyme system used
electrochemistry biosensor–glucose dehydrogenase
Price per disposable reagent system unit
pricing based on volume contingent on contract price
No. of dispos. reag. system units per basic package
50 strips per vial and 100 per box 50 strips per vial
No. of times analyses performed using 1 reag. system unit
1 1
Dispos. units shelf life/Reag. unit storage requirements
24 months from date of manufacture/none 18 months, stable until expiration on vial/<90°F, do not freeze
Digital readout size/Keypad input capability
varies and is defined by the particular field/numeric, alphabetic 4 lines by 20 characters LCD/menu selection, numeric
How results are displayed
true values true values
Specimen types/Sampling techniques
whole blood/drop (arterial, venous, capillary, neonatal) whole blood/arterial, venous, capillary, neonate (including cord blood)
Minimum specimen volume required
1.2 µL 4 µL
Suitable for samples from well/Sick neonates
yes/yes yes/yes
Time from sample intro. to result availability
6 seconds 26 seconds
Batteries used/No. used/Avg. life of one set
3.7 Li Polymer (rechargeable/replaceable)/1/24–36 months 3 V lithium/2/~700 tests
Avg. expected life of device/Mean time between failures
5+ years/— 5 years/10,000 tests
Device warranty/Service options
2 years (extended 5-year warranty at additional cost)/meter replacement AccuData GTS Plus/GTS system will be free from defects in materials &
workmanship through life of Accu-Chek Comfort Curve test strip contract;
overnight replacement, according to warranty policy, is available 24/7 365
days per year
Loaners provided
yes replaced under warranty
User list or user group
no yes (contact local account manager)
Toll-free No. for customer questions/Hours
800-458-5813/24 hours, 7 days, all year 800-440-3638/24 hours, 365 days per year
Training and certif. program/No. training days provided
yes/defined during implementation planning yes/site-specific according to No. of employees
Avg. time for lab to complete maintenance
no user maintenance none
Special cleansing procedures
no no
Internal QC recommended or required
CLIA requirements 2 levels per day daily, 2 levels
Between instrument CV (based on PT) at these levels:
• <50 mg/dL
53.8 mg/dL SD=4.1 (6,088 labs)
• 100–200 mg/dL
191.4 mg/dL CV=4.7% (3,096 labs)
• >400 mg/dL
228.5 mg/dL CV=4.6% (6,099 labs)
• Program name, year/Challenge No./Level of mean glucose challenge sample
CAP, 2001/WBG-C/see above
Accuracy/Compared to what reference method or device
R2=0.9978, slope=1.0127–2.0975/YSI 2300 y=0.991 x + 8.4, r=0.980/glucose hexokinase-Hitachi
Precision/Compared to what reference method or device
within run (whole blood=1.9%–3.6%) & (day to day=3.4%–4.7%) linearity
standards/—
controls: low SD=2.83 mg/dL, mid CV=3.08%, high CV=2.82%; blood: low
SD=1.5 mg/dL, mid CV=3.2%, high CV=3.2%/glucose hexokinase
Linear range
10–600 mg/dL 10–600 mg/dL
Suggested dynamic, measurement range
10–600 mg/dL 10–600 mg/dL
Contraindications
per labeling
Known interferences/High-altitude interference
none/no, operates at altitudes up to 15,000 feet per labeling/none up to 10,150 feet
Restrictions based on hematocrit
none (no Hct interference) yes, glucose <200 mg/dL, 20%–65%; glucose >200, 20%–55%
Electronic, optical function checks
electronic checks for out-of-range glucose results, dosing, out-of-range
Hct results
meter cradle communication with Advantage meter, GTS with code key,
battery voltage test, internal database memory check, internal
configuration check
Sample quantity checks
RapidFill sampling electronically checks for correct strip dosing built-in electronic strip check, visual confirmation of sample volume
When auto lock or shutdown occurs
options include user ID failure, QC failure, required docking for data
transfer
user ID failure (valid op.), QC failure, patient ID length, incorrect code key,
incorrect Advantage meter
User defines QC lockout intervals/Lockout can be circumvented
yes/no, not if configured yes/yes (information management system identifies operators who violate
hospital policy)
Device supports bar-code scanning of
operator & patient identifiers, reagent, lot No., QC lots operator & patient identifiers, comment codes
Method of analyst ID/ID required
medical record ID No., medical billing ID No., Accession ID No./ID required numeric input or bar-code wand scan/yes
Internal memory size/Max. No. patient results stored
1,000 patient samples, 200 QC samples, 4,000 operators/1,000 tests 1,000 total patient, control, linearity, proficiency tests/1,000
Information transfer capability:
• Meters connect to
Instrument Manager (NovaNet or Laboratory Data Systems AegisPOC) to
Data Manager (Telcor QML/Quick-Linc or AegisPOC) then to LIS if required
information management system, which in turn connects to LIS/HIS
How meters are connected to external system to upload results/
No. installations
hospital network/—; wireless tote/— direct serial/—, modem dial-in/—, hospital network/—
• Info. contained in transmission to external system
device unique identifier, operator & patient IDs, results, QC identifiers device unique identifiers, operator & patient IDs, results, QC identifiers,
strip lot Nos., download location, comment codes, proficiency & linearity
samples
Hardware/software for data mgmt. system
connects to Telcor QML and Laboratory Data Systems AegisPOC MAS RALS-Plus, MAS RALS-Lite†, MAS RALS-Notebook
No. of different mgmt. reports system can produce
provided by Telcor and Laboratory Data Systems varies by Data Manager (customer defined)
Contents downloaded from DMS to meter
strip lot numbers, valid control values, valid operator IDs, patient
demographics, configuration files, physician IDs, diagnostic codes
strip & QC lot Nos., valid operator IDs, valid control values, linearity values
System connected (live installations) to which LISs/HISs:
• using screen animation/screen scraping
available through Telcor & Laboratory Data Systems all major LIS vendors including Cerner, Misys, McKesson, Meditech,
SoftLab, Siemens, SIA Molis, Opus, others**
• using standard HL7 interface
yes
• using proprietary protocol interface
no
Use 3rd-party interfacing tool/engine for LIS/HIS interfaces
yes, Telcor QML/Quick-Linc, Laboratory Data Systems AegisPOC MAS
Distinguishing features (provided by vendors)
measures and eliminates interferences from hematocrit, oxygen,
acetaminophen, ascorbic acid, uric acid, and other electrochemical
substances; no interference from maltose, galactose, or xylose; no
calibration codes required; results reported in 6 seconds using 1.2 µL
of sample; unlimited manual test entry; 1-D and 2-D bar-code capability
available
proven bi-directional network connection from AccuData GTS/GTS
Plus to LIS/HIS; ADT data interface with RALS-Plus/DataCare POC;
uses the Accu-Chek Comfort Curve test strip; universal sampling due
to oxygen-independent chemistry, with reliable results at varying
hematocrit levels
*combined AccuData GTS and AccuData GTS Plus sales
Roche exclusive
**both scripted/HL7 are available
Tabulation does not represent an endorsement by the College of American Pathologists.